{"product_id":"pq3198-94","title":"PQ3198-94","description":"\u003ch4 class=\"hdg2-04\"\u003eBasic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)\u003c\/h4\u003e\n\u003cdiv id=\"tablefield-wrapper-paragraph-32378-field_table_spec-0\" class=\"tablefield-wrapper\"\u003e\n\u003cdiv class=\"table-scroll-01\"\u003e\n\u003cdiv class=\"inner\"\u003e\n\u003ctable class=\"table-01\" style=\"width: 99.9589%; height: 1075.5px;\"\u003e\n\u003ctbody\u003e\n\u003ctr class=\"odd\" style=\"height: 58.75px;\"\u003e\n\u003cth class=\"row_0 col_0\" style=\"width: 18.3505%; height: 58.75px;\"\u003eMeasurement line type\u003c\/th\u003e\n\u003ctd class=\"row_0 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 58.75px;\"\u003eSingle-phase 2-wire, Single-phase 3-wire, Three-phase 3-wire or Three-phase 4-wire plus one extra input channel for voltage, current, power measurement (AC or DC measurement)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 39.1667px;\"\u003e\n\u003cth class=\"row_1 col_0\" style=\"width: 18.3505%; height: 39.1667px;\"\u003eVoltage ranges\u003c\/th\u003e\n\u003ctd class=\"row_1 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 39.1667px;\"\u003eVoltage measurement: 600.00 V rms\u003cbr\u003eTransient measurement 6.0000 kV peak\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"odd\" style=\"height: 19.5833px;\"\u003e\n\u003cth class=\"row_2 col_0\" style=\"width: 18.3505%; height: 19.5833px;\"\u003eCurrent ranges\u003c\/th\u003e\n\u003ctd class=\"row_2 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 19.5833px;\"\u003e500.00 mA to 5.0000 kA AC (depends on current sensor in use)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 39.1667px;\"\u003e\n\u003cth class=\"row_3 col_0\" style=\"width: 18.3505%; height: 39.1667px;\"\u003ePower ranges\u003c\/th\u003e\n\u003ctd class=\"row_3 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 39.1667px;\"\u003e300.00 W to 3.0000 MW (determined automatically based on voltage and current range in use)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"odd\" style=\"height: 58.75px;\"\u003e\n\u003cth class=\"row_4 col_0\" style=\"width: 18.3505%; height: 58.75px;\"\u003eBasic accuracy\u003c\/th\u003e\n\u003ctd class=\"row_4 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 58.75px;\"\u003eVoltage: ±0.1% of nominal voltage\u003cbr\u003eCurrent: ±0.1 % rdg. ±0.1 % f.s. + current sensor accuracy\u003cbr\u003eActive power: ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 528.75px;\"\u003e\n\u003cth class=\"row_5 col_0\" style=\"width: 18.3505%; height: 528.75px;\"\u003eMeasurement items\u003c\/th\u003e\n\u003ctd class=\"row_5 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 528.75px;\"\u003e1. Transient voltage : 2 MHz sampling\u003cbr\u003e2. Frequency cycle : Calculated as one cycle, 40 to 70 Hz\u003cbr\u003e3. Voltage (1\/2) RMS: one cycle calculation refreshed every half cycle\u003cbr\u003eCurrent (1\/2) RMS: half-cycle calculation\u003cbr\u003e4. Voltage swell, Voltage dips, Voltage interruption\u003cbr\u003e5. Inrush current\u003cbr\u003e6. Voltage waveform comparison\u003cbr\u003e7. Instantaneous flicker value: As per IEC61000-4-15\u003cbr\u003e8. 200 ms frequency: Calculated as 10 or 12 cycles,40 to 70 Hz\u003cbr\u003e9. 10 sec frequency: Calculated as the whole-cycle time during the specified 10 s period, 40 to 70 Hz\u003cbr\u003e10. Voltage waveform peak, Current waveform peak\u003cbr\u003e11. Voltage, Current, Active power, Apparent power, Reactive power, Active energy, Reactive energy, Power factor, Displacement power factor, Voltage unbalance factor, Current unbalance factor, and efficiency\u003cbr\u003e12. High-order harmonic* component (voltage\/ current): 2 kHz to 80 kHz\u003cbr\u003e*Supraharmonic\u003cbr\u003e13. Harmonic\/ Harmonic phase angle (voltage\/ current), Harmonic power: 0th to 50 th orders\u003cbr\u003e14. Harmonic voltage-current phase angle: 1th to 50 th orders\u003cbr\u003e15. Total harmonic distortion factor (voltage\/ current)\u003cbr\u003e16. Inter harmonic (voltage\/ current): 0.5 th to 49.5 th order\u003cbr\u003e17. K Factor (multiplication factor)\u003cbr\u003e18. IEC Flicker, Δ V10 Flicker\u003cbr\u003e19. Mains signaling voltage\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"odd\" style=\"height: 58.75px;\"\u003e\n\u003cth class=\"row_6 col_0\" style=\"width: 18.3505%; height: 58.75px;\"\u003eRecord\u003c\/th\u003e\n\u003ctd class=\"row_6 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 58.75px;\"\u003eRepeated ON: 1 year, Maximum recording event: 9999 × 366 days (up to 9999 events per day)\u003cbr\u003eRepeated off: 35 days, maximum recording event: 9999 events\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 39.1667px;\"\u003e\n\u003cth class=\"row_7 col_0\" style=\"width: 18.3505%; height: 39.1667px;\"\u003eInterfaces\u003c\/th\u003e\n\u003ctd class=\"row_7 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 39.1667px;\"\u003eSD\/SDHC memory card, LAN (HTTP server function \/ FTP function), USB2.0 (for communication)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"odd\" style=\"height: 19.5833px;\"\u003e\n\u003cth class=\"row_8 col_0\" style=\"width: 18.3505%; height: 19.5833px;\"\u003eDisplay\u003c\/th\u003e\n\u003ctd class=\"row_8 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 19.5833px;\"\u003e6.5-inch TFT color LCD (640 × 480 dots)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 58.75px;\"\u003e\n\u003cth class=\"row_9 col_0\" style=\"width: 18.3505%; height: 58.75px;\"\u003ePower supply\u003c\/th\u003e\n\u003ctd class=\"row_9 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 58.75px;\"\u003eAC adapter Z1002 (100 V to 240 V AC, 50\/60 Hz, rated current 1.7 A), Battery Pack Z1003 (Continuous use: 180 minutes, Charging time: Max. 5 hr 30 m with AC adapter)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"odd\" style=\"height: 57.1667px;\"\u003e\n\u003cth class=\"row_10 col_0\" style=\"width: 18.3505%; height: 57.1667px;\"\u003eDimensions and mass\u003c\/th\u003e\n\u003ctd class=\"row_10 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 57.1667px;\"\u003e300 mm (11.81 in)W × 211 mm (8.31 in)H × 68 mm (2.68 in)D, 2.6 kg (91.7 oz) (including Battery Pack Z1003)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr class=\"even\" style=\"height: 97.9167px;\"\u003e\n\u003cth class=\"row_11 col_0\" style=\"width: 18.3505%; height: 97.9167px;\"\u003eIncluded accessories\u003c\/th\u003e\n\u003ctd class=\"row_11 col_1\" colspan=\"6\" style=\"width: 79.9283%; height: 97.9167px;\"\u003eInstruction manual ×1, Measurement guide ×1, Voltage Cord L1000 ×1 set (Red\/ Yellow\/ Blue\/ Gray each 1, Black 4, 3m (9.84ft) length, Alligator clip ×8), Spiral tube x20, Color clip, AC Adapter Z1002 ×1, Strap ×1, USB cable (1 m 3.28 ft length) ×1, Battery pack Z1003 ×1, SD Memory Card 2GB Z4001 ×1, Flexible Current Sensors CT7045 (4 units)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e","brand":"Hioki-nz","offers":[{"title":"Default Title","offer_id":47239594934443,"sku":null,"price":17960.0,"currency_code":"NZD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0765\/7789\/8667\/files\/PQ3198EN.jpg?v=1781661511","url":"https:\/\/hioki.nz\/products\/pq3198-94","provider":"Hioki NZ","version":"1.0","type":"link"}